Real-time weather monitor is not working / will not start [Article]
The real-time weather monitor displays current conditions (updated every 15 seconds). It requires a direct wired connection or direct wireless connection from the weather statio
Is the pH pro waterproof? (2100A) [Article]
No, the pH pro is only splashproof. Do not make any measurements by submerging the meter completely in the sample solution. If the meter is inadvertently dropped into
My Cardy sensor will not hold calibration (2300) [Article]
These sensors wear out over time. Replace sensor. Item # 2340, or Contact Spectrum.
Cardy NO3- Trouble Shooting
What are the correct standards?
The IQ 150 meter calibrates very slowly (2162) [Article]
The calibration of the IQ 150 can take up to one and a half minutes for each calibrating solution (pH 7 and pH 4). The probe should be hydrated by soaking in water for several m
Error Message: No Carrier Found [Article]
The telephone modem at the weather station has answered but cannot establish a synchronous connection with the computer modem. One possible cause is that the baud
Error Message: SpecWare trying to perform a Fast Launch [Article]
When communicating with remotely located WatchDog weather stations via radio transceivers or a telephone connection, it is not possible to bring up the Launch Options screen to modify th
How can I test the life left in my pH probe? [Article]
A simple test can be performed using Windex glass cleaner with ammonia and any type of soft drink. The procedure is as follows: Turn your meter on and place the probe in a so
Does the pH meter need to be calibrated before each use? [Article]
No, most pH meters hold calibration very well. If you are using your meter every day then check it once a week. If you get a reading that is above or below what you were expe
Error message on pH Pro (2100A) [Article]
Indicates calibration error. This can be due to:
-No standard solution on the sensor
-Bubbles on the surface of the ISFE