  1. No GPS Found with Field Scout TDR MeterNo GPS Found with Field Scout TDR Meter [Article]


  2. Start Depth > 0! messageStart Depth > 0! message [Article]
    The meter cannot begin taking readings until the probe senses the tip is at the soil surface. The meter can sense a false non-zero depth if the user’s feet are too close to the shaft, if

  3. Meter prematurely gives "READING COMPLETE" message.Meter prematurely gives "READING COMPLETE" message. [Article]
    Ultrasonic depth sensor is incorrectly sensing that the meter is being withdrawn from the soil. This is similar to second part of solution for !! Depth Error !! error mes

  4. READINGS TOO HIGH OR TOO LOWREADINGS TOO HIGH OR TOO LOW [Article]
    Calibration may be off. The calibration procedure is published in the

  5. !Error- Too Fast! message!Error- Too Fast! message [Article]
    Probe is being inserted too quickly. If the meter is inserted faster than 1 inch per 1 second, the ASABE standard states that the reading s

  6. !! Depth Error !! message!! Depth Error !! message [Article]
    This error message indicates that the meter missed a depth reading. This can happen when: a) The speed of insertion exceeds the meter's abilit

  7. SC900 meter will not turn on after installing new batteries.SC900 meter will not turn on after installing new batteries. [Article]
    Pull the batteries and wait 3 or more minutes before inserting them back into the meter. This will allow the cap

  8. LCD not workingLCD not working [Article]
    Battery power not enough. Remove four screws on meter's front plates and replace batteries.

  9. DELETE button not functioning on SC900DELETE button not functioning on SC900 [Article]
    ...abled. To enable the data logger, connect the SC900 meter to Field Scout software and bring up the Meter Settings screen.&nb

  10. Reading in Relative Water Content Mode.Reading in Relative Water Content Mode. [Article]
    The TDR meters have 2 modes of operation, volumetric water content (VWC) and relative water content (RWC). In RWC mode, soil moisture is reported on a scale of 0 - 100 with the low and high mo

  11. Is the pH pro waterproof? (2100A)Is the pH pro waterproof? (2100A) [Article]
    No, the pH pro is only splashproof. Do not make any measurements by submerging the meter completely in the sample solution. If the meter is inadvertently dropped into

  12. What are the Cardy Twin pH's specifications? (2103)What are the Cardy Twin pH's specifications? (2103) [Article]
    Range: pH 2.00 to 12.00; Accuracy/Repeatability: +/-0.1pH Sensor Lifespan: 150 - 300 s

  13. Cardy Twin pH meter does not maintain calibration (2103)Cardy Twin pH meter does not maintain calibration (2103) [Article]
    - Make sure it turns on OK - Get both solutions pH 7 (yellow) and pH 4 (pink). Connect the 2 dots in an egg shape with a few drops.

  14. Cardy Twin Calibration fails / Calibration light continues blinking after calibration effort (2103)Cardy Twin Calibration fails / Calibration light continues blinking after calibr... [Article]
    - Inspect sensor in bright light for cracks or flaws. - Place 4.0 standard solution on sensor and attempt reading. Fa

  15. Cardy Twin pH Meter is giving no (or incorrect) reading (2103)Cardy Twin pH Meter is giving no (or incorrect) reading (2103) [Article]
    - Inspect sensor in bright light for cracks or flaws. - Place 4.0 standard solution on sensor and attempt reading. Fa

  16. My readings are off. (2205)My readings are off. (2205) [Article]
    When was the last time you calibrated the meter? I would suggest you Re-calibrate and take a reading again. How old is

  17. Can I directly insert the meter into either soil or liquid media? (2245)Can I directly insert the meter into either soil or liquid media? (2245) [Article]
    The meter is not designed to get direct readings from media. With this unit you should still prepare the media in a cup and then take a sample and place it in th

  18. I used regular water when calibrating and the meter is not calibrating correctly. (2245)I used regular water when calibrating and the meter is not calibrating correctly... [Article]
    We suggest you use deionized water.

  19. I took several readings from pots and after about 15 or so I think the probe was (2265FS)I took several readings from pots and after about 15 or so I think the probe was... [Article]
    Try wipping off exccess media from the probe. You may need to re-calibrate the meter again. We recommend you calibrate it once a month it should take y

  20. How do I calibrate the meter? (2300)How do I calibrate the meter? (2300) [Article]
    ...(slope). Please refer to calibration procedure in Cardy Nitrate manual which can be downloaded on ou...